Description
The filling solution should be filled into the electrode with designated plastic bottle to avoid damage the membrane. The level of the filling solution should reach 2/3 height of the electrode inner chamber. Then fasten the electrode gap, shake the electrode till the air bubble goes out, dry the electrode body and install to the designated position. A new electrode can be put in use immediately. Should it be unstable, please use fresh plasma to activate the electrodes till it reached stable status, it may take approximately 30-60 minutes.
Packing List
1801 K Electrode | 1pc |
1802 Na Electrode | 1pc |
1803 Cl Electrode | 1pc |
1804 Ca Electrode | 1pc |
1805 pH Electrode | 1pc |
1807 Ref Electrode | 1pc |
Features:
1. Rinsing Method: quickly rinsing solenoid valve, cleaning thoroughly electrode pipe;
2. Measuring Time: less than 50S (depends on electrode in stability);
3. Display: 5.6 inch Color Display touch screen;
4. Sensor: intelligent Sampling system, full automatic taking sample;
5. Parameters: K+, Na+, Cl-,Ca++,pH ,Li and Ref;
6. Temperature: 5-40 ℃, less than 85% relative humidity
7. Power Requirements: 220V±22V,50Hz±1Hz
Product Specifications:
Electrode | Range | Resolution | CV |
PH Hydrogen ion selective, flow-through electrode | 4-9mmol/L | 0.01mmo/L | <1% |
K potassium ion selective,flow-through electrode | 0.5-10mmol/L | 0.01mmo/L | <1% |
Na Sodium ion selective, flow-through electrode | 80-200mmol/L | 0.1mmo/L | <1% |
CL chloride ion selective, flow-through electrode | 50-200mmol/L | 0.1mmo/L | <1% |
CA calcium ion selective, flow-through electrode | 0.1-5mmol/L | 0.01mmo/L | <2% |
Li lithium ion selective, flow-through electrode | 0.3-3mmo/L | 0.01mmo/L | <2% |